From: A sensitive method for analyzing fluconazole in extremely small volumes of neonatal serum
| Conditions | Concentration (μg/mL) | Accuracy (%) | Precision (%) |
|---|---|---|---|
| Intraday | 0.01 | −4.6 | 3.1 |
| 0.1 | 4.7 | 4.6 | |
| 1.0 | −0.4 | 2.1 | |
| 10.0 | 7.4 | 3.4 | |
| 50.0 | 5.0 | 6.0 | |
| 100.0 | −7.7 | 5.8 | |
| Interday | 0.01 | −4.4 | 4.3 |
| 0.1 | 0.8 | 2.7 | |
| 1.0 | 3.0 | 0.3 | |
| 10.0 | 3.2 | 2.5 | |
| 50.0 | −2.5 | 2.5 | |
| 100.0 | −0.3 | 5.4 |